Ultra Mouse Insulin ELISA (EIA) Kit

      Catalog#
         90080
 Ultra Sensitive Mouse Insulin ELISA Kit
 
                 Low and High Range Mouse Insulin Assay Using the Same Kit
 
Sample Size : 5µl



Samples:

Serum, plasma, or fluid
Tests: 96 wells (8 wells x 12 modules)
Reagents: In liquid form (except standard)
Assay Range: Low Range Assay: 0.1 - 6.4 ng/ml
Wide Range Assay: 0.1 - 12.8 ng/ml*
High Range Assay: 1 - 64 ng/ml


* Intended for screening purpose

Assay Time: Same day procedure
Precision: Intra-assay precision CV ≤ 10.0%
Inter-assay precision CV ≤ 10.0%

Recovery: When mouse insulin was spiked to mouse serum, the recovery of mouse insulin was 100% ± 15%
Specificity:
Substance Reactivity
Mouse Insulin 100%
Rat Insulin 100%*
Hamster Insulin Detected*
Rat C-peptide Not detected
Rat pancreatic polypeptide Not detected
Porcine glucagon Not detected
Human Insulin like growth factor-I Not detected
Human Insulin like growth factor-II Not detected

*Can vary from lot to lot. Specific cross reactivity data is included with each kit

Typical Standard Curve:
Highlights: Small Sample Volume: Only 5µl

High Sensitivity: 100 pg/ml using 5µl sample and 5 pg/ml using 100µl sample

Low and High Insulin Assay:Both using the same kit

Precision:
Intra-assay and inter-assay precision CV ≤ 10.0%

Protocol: Same protocol for serum, plasma, or fluid assay

Effect of Hemolysis: Almost no effect of hemolysis tested up to 20mg/ml.

Summary of Assay
Ultra Sensitive Mouse Insulin ELISA Kit (Catalog# 90080)
                                         Low/Wide and High Range Assay
 
 
Affix the antibody-coated microplate module on the frame
Dispense 95µl sample diluent per well
Pipette 5µl of the sample (or working mouse insulin standard**) to the well
Incubate the microplate for 2 hours at 4°C.
Wash the well five times with washing buffer
Dispense 100µl anti-insulin enzyme conjugate per well
Incubate microplate for 30 minutes at room temperature
Wash each well seven times with washing buffer
Dispense 100µl enzyme substrate solution per well
Incubate microplate at room temperature while avoiding exposure to light
                                            
     40 minutes-                                10 minutes-
low/wide range assay                         high range assay
                                            
Stop the reaction by adding 100µl Enzyme reaction stop solution per well
Measure the absorbance at A 450nm and subtract A630nm values within 30 minutes
Calculate the mouse insulin concentration using the standard curve
 
            *Each well should be washed with 300µl of wash buffer. Aspirate the wells completely so all                   
excess solution is removed.                                                                                                  
 
         ** Preparation of standards are different for both low and high range assay. Please refer to the                     
             product insert for more details
 
Crystal Chem Inc.        Tel: (630) 889-9003        Fax: (630) 889-9021        E-mail: sales@crystalchem.com