home
|
email
Home
Products
Inquiries
Ordering Info
About Us
Contact Us
Featured Products
Coenzyme A & Derivatives
Ultra Sensitive Mouse Insulin ELISA Kit
Rat Leptin ELISA Kit
dNTP Solution
Anti-Calcitonin Mono
Nucleotides
Please
contact us
for more info...
Ultra Mouse Insulin ELISA (EIA) Kit
Catalog#
90080
Ultra Sensitive Mouse Insulin ELISA Kit
Low and High Range Mouse Insulin Assay Using the Same Kit
Sample Size :
5µl
Samples:
Serum, plasma, or fluid
Tests:
96 wells (8 wells x 12 modules)
Reagents:
In liquid form (except standard)
Assay Range:
Low Range Assay: 0.1 - 6.4 ng/ml
Wide Range Assay: 0.1 - 12.8 ng/ml*
High Range Assay: 1 - 64 ng/ml
* Intended for screening purpose
Assay Time:
Same day procedure
Precision:
Intra-assay precision CV ≤ 10.0%
Inter-assay precision CV ≤ 10.0%
Recovery:
When mouse insulin was spiked to mouse serum, the recovery of mouse insulin was 100% ± 15%
Specificity:
Substance
Reactivity
Mouse Insulin
100%
Rat Insulin
100%*
Hamster Insulin
Detected*
Rat C-peptide
Not detected
Rat pancreatic polypeptide
Not detected
Porcine glucagon
Not detected
Human Insulin like growth factor-I
Not detected
Human Insulin like growth factor-II
Not detected
*Can vary from lot to lot. Specific cross reactivity data is included with each kit
Typical Standard Curve:
Highlights:
Small Sample Volume:
Only 5µl
High Sensitivity:
100 pg/ml using 5µl sample and 5 pg/ml using 100µl sample
Low and High Insulin Assay:
Both using the same kit
Precision:
Intra-assay and inter-assay precision CV ≤ 10.0%
Protocol:
Same protocol for serum, plasma, or fluid assay
Effect of Hemolysis:
Almost no effect of hemolysis tested up to 20mg/ml.
Summary of Assay
Ultra Sensitive Mouse Insulin ELISA Kit (Catalog# 90080)
Low/Wide and High Range Assay
Affix the antibody-coated microplate module on the frame
Dispense 95µl sample diluent per well
Pipette 5µl of the sample (or working mouse insulin standard
**
) to the well
Incubate the microplate for 2 hours at 4°C.
Wash the well five times with washing buffer
Dispense 100µl anti-insulin enzyme conjugate per well
Incubate microplate for 30 minutes at room temperature
Wash each well seven times with washing buffer
Dispense 100µl enzyme substrate solution per well
Incubate microplate at room temperature while avoiding exposure to light
40 minutes-
10 minutes-
low/wide range assay high range assay
Stop the reaction by adding 100µl Enzyme reaction stop solution per well
Measure the absorbance at A
450nm
and subtract A
630nm
values within 30 minutes
Calculate the mouse insulin concentration using the standard curve
*Each well should be washed with 300µl of wash buffer. Aspirate the wells completely so all
excess solution is removed.
** Preparation of standards are different for both low and high range assay. Please refer to the
product insert for more details
home
|
products
|
inquiries
|
ordering info
|
about us
|
contact us
|
terms and conditions
Crystal Chem Inc. Tel: (630) 889-9003 Fax: (630) 889-9021 E-mail:
sales@crystalchem.com